GHB and Related Compounds in Human Blood by LC-MS/MS

PeakstR (min)Precursor ionProduct ion
1.γ-Hydroxybutyric acid (GHB)1.42105.287.0
2.1,4-Butanediol (1,4-BD)1.4291.055.0
3.γ-Butyrolactone (GBL)1.7587.045.0
image
LC_CF0732
ColumnForce C18 (cat.# 963431E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:3 µm
Pore Size:100 Å
Guard Column:Force C18 EXP guard column cartridge 5 mm, 3.0 mm ID, 3 µm (cat.# 963450253)
Temp.:30 °C
Sample
Diluent:Water
Conc.:500 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:0.5% Formic acid in water
B:0.5% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.7955
0.500.7955
3.000.75050
3.010.7955
5.000.7955
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC
Notes100 μL of whole human blood was fortified at 50 μg/mL with GHB, GBL, 1,4-BD, and GHB-D6 (IS) using 5 μL of 1 mg/mL solutions. The blood was precipitated with 380 μL methanol. The sample was then vortexed at 1000 rpm for 10 seconds and centrifuged at 3000 rpm for 10 minutes at 10 °C. 50 μL of the supernatant was removed and diluted to 1 mL with water. The sample was then vortexed and subjected to LC-MS/MS analysis. (Internal standard not shown on chromatogram.)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

GHB

,

GBL

,

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us