This chromatogram may contain analytes that are synonyms of your search terms.
EPA Method 610 Polycyclic Aromatic Hydrocarbons on Pinnacle DB PAH
Peaks | |
---|---|
1. | Naphthalene![]() |
2. | Acenaphthylene![]() |
3. | 1-Methylnaphthalene![]() |
4. | 2-Methylnaphthalene![]() |
5. | Acenaphthene![]() |
6. | Fluorene![]() |
7. | Phenanthrene![]() |
8. | Anthracene![]() |
9. | Fluoranthene![]() |
Peaks | |
---|---|
10. | Pyrene![]() |
11. | Benzo[a]anthracene![]() |
12. | Chrysene![]() |
13. | Benzo[b]fluoranthene![]() |
14. | Benzo[k]fluoranthene![]() |
15. | Benzo[a]pyrene![]() |
16. | Dibenzo[a,h]anthracene![]() |
17. | Benzo[ghi]perylene![]() |
18. | Indeno[1,2,3-cd]pyrene![]() |

LC_EV0469
Column | Pinnacle DB PAH (cat.# 9470252) | ||||||||||
Dimensions: | 50 mm x 2.1 mm ID | ||||||||||
Particle Size: | 1.9 µm | ||||||||||
Pore Size: | 140 Å | ||||||||||
Temp.: | 30 °C | ||||||||||
Sample | |||||||||||
Diluent: | Acetonitrile | ||||||||||
Conc.: | 20 µg/mL each component | ||||||||||
Inj. Vol.: | 2 µL | ||||||||||
Mobile Phase | |||||||||||
A: | Water | ||||||||||
B: | Acetonitrile | ||||||||||
| |||||||||||
Flow: | 0.6 mL/min | ||||||||||
Detector | UV/Vis @ 254 nm | ||||||||||
Instrument | Jasco X-LC |