share:

EPA 8310 PAH Mix on Pinnacle DB PAH

Peaks
1.Naphthalene
2.Acenaphthylene
3.1-Methylnaphthalene
4.2-Methylnaphthalene
5.Acenaphthene
6.Fluorene
7.Phenanthrene
8.Anthracene
9.Fluoranthene
Peaks
10.Pyrene
11.Benzo[a]anthracene
12.Chrysene
13.Benzo[b]fluoranthene
14.Benzo[k]fluoranthene
15.Benzo[a]pyrene
16.Dibenzo[a,h]anthracene
17.Benzo[ghi]perylene
18.Indeno[1,2,3-cd]pyrene
image
LC_EV0521
ColumnPinnacle DB PAH (cat.# 9470252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.9 µm
Pore Size:140 Å
Temp.:30 °C
SampleEPA Method 8310 PAH Mixture (cat.# 31841)
Diluent:acetonitrile
Conc.:10 µg/mL
Inj. Vol.:1 µL
Mobile Phase
A:water
B:acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.86040
20.84060
40.80100
4.50.80100
4.510.86040
50.86040
Max Pressure:724 bar
DetectorPhoto diode array @ 254, 4.8 nm
InstrumentWaters
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Pinnacle

,

Pinnacle DB

,

UPLC

,

polynuclear aromatic hydrocarbons

,

polycyclic aromatic hydrocarbons

,

PAHs

,

UHPLC

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.