share:

This chromatogram may contain analytes that are synonyms of your search terms.

Explosives on Raptor ARC-18 by EPA Method 8330B

Peaks
1.HMX
2.RDX
3.1,3,5-Trinitrobenzene
4.1,4-Dinitrobenzene
5.1,3-Dinitrobenzene
6.3,5-Dinitroaniline
7.Nitrobenzene
8.Nitroglycerin
9.Tetryl
Peaks
10.2,4,6-Trinitrotoluene
11.2-Amino-4,6-dinitrotoluene
12.4-Amino-2,6-dinitrotoluene
13.2,4-Dinitrotoluene
14.2,6-Dinitrotoluene
15.2-Nitrotoluene
16.4-Nitrotoluene
17.3-Nitrotoluene
18.PETN
image
LC_EV0530
ColumnRaptor ARC-18 (cat.# 9314A1E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:2.7 µm
Temp.:35 °C
Sample8330B Nitroaromatics and nitramine mix (cat.# 33204)
1,4-Dinitrobenzene (cat.# 33205)
Diluent:Water
Conc.:20 µg/mL
Inj. Vol.:3 µL
Mobile Phase
A:Water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.001.07525
22.51.07525
DetectorUV/Vis @ 210/254 nm
Cell Temp:35 °C
Flow Cell Size:Semi-micro
InstrumentShimadzu UFLCXR
NotesThis is the recommended confirmation column for a two-column analysis. See LC_EV0531 for the primary column.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

8330B

,

explosives

,

ARC-18

,

,

Raptor

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us