share:

This chromatogram may contain analytes that are synonyms of your search terms.

Pharmaceuticals and Personal Care Products (PPCPs) on Raptor Biphenyl (1.8 μm)

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonProduct IonPolarity
1.Acetaminophen0.5910151.8110.265.1+
2.Lincomycin0.905407.4126.3359.3+
3.Cotinine1.0850176.980.298.2+
4.Codeine1.2250300.5165.2215.2+
5.Trimethoprim1.4110290.9230.2261.2+
6.Sulfadiazine1.5350251.4156.292.2+
7.Norfloxacin1.6250320.4231.1302.2+
8.Sulfathiazole1.6750256.3156.392.2+
9.Ampicillin1.76100382.4106.2223.3+
10.Ofloxacin1.8350362.4318.2261.2+
11.Ciprofloxacin1.8910332.3231.1314.2+
12.Oxytetracycline1.9050461.2426.2443.2+
13.Sulfamerazine2.0150265.3108.292.3+
14.Tetracycline2.1150445.4410.2427.2+
15.Caffeine2.1610195.2138.2110.2+
16.Sulfamethizole2.2550271.4156.292.1+
17.Sulfamethoxazole2.3310253.9156.1108.2+
18.Sulfamethazine2.455279.3186.2124.2+
19.Sulfachloropyridazine2.4950285.2156.1108.1+
20.Clinafloxacin2.5850366.3305.1348.1+
21.Azithromycin2.60100749.5591.5158.2+
22.Carbadox2.6450263.3231.2129.2+
23.Chlortetracycline2.81100479.2444.1462.2+
24.Diphenhydramine2.895256.4167.3165.3+
25.Sulfadimethoxine3.045311.3156.3108.2+
26.Fluoxetine3.1910310.344.2148.3+
27.Erythromycin3.3210734.7158.2576.4+
28.Penicillin G3.3850367.4160.2217.2+
29.Diltiazem3.555415.4178.3109.2+
30.Clarithromycin3.625748.6158.2590.5+
31.Carbamazepine3.7010237.3194.2193.2+
32.Oxolinic acid3.7150262.3216.1244.1+
33.Tylosin3.72100948.6174.3156.3+
34.Penicillin V3.7350383.3160.1114.2+
35.Oxacillin3.9550434.2160.2144.2+
36.Naproxen4.1450229.3169.3170.2-
37.Digoxin4.1650839.5779.5649.5-
38.Flumequine4.275262.3202.2244.2+
39.Triclosan4.3350287.135.0--
40.Warfarin4.375309.2163.2251.2+
41.Gemfibrozil4.7150249.3121.2--
image
LC_EV0539
ColumnRaptor Biphenyl (cat.# 9309252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Pore Size:90 Å
Temp.:40 °C
SamplePharmaceuticals mix #1 (cat.# 31116)
Pharmaceuticals mix #2 (cat.# 31118)
Compounds not present in these mixes were obtained separately.
Diluent:0.1% Acetic acid in water
Conc.:5-100 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Acetic acid in water
B:0.1% Acetic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.58515
5.000.51090
5.010.58515
7.000.58515
DetectorMS/MS
Mode:MRM
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PPCP

,

1694

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us