share:

Perfluorinated Compounds (PFCs) on Raptor C18 2.7 μm

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Perfluorobutanoic acid (PFBA)0.885213.1169.1
2.Perfluoropentanoic acid (PFPeA)1.815263.1219.1
3.Perfluorobutane sulfonic acid (PFBS)2.075299.180.0
4.1H, 1H, 2H, 2H-Perfluorohexane sulfonic acid (4:2 FTS)2.615327.1307.1
5.Perfluorohexanoic acid (PFHxA)2.695313.1269.1
6.Perfluoropentane sulfonic acid (PFPeS)2.825349.180.0
7.Perfluoroheptanoic acid (PFHpA)3.315363.2319.1
8.Perfluorohexane sulfonic acid (PFHxS)3.375399.180.0
9.2-Perfluorohexyl ethanoic acid (FHEA)3.43250377.1293.1
10.1H, 1H, 2H, 2H-Perfluorooctane sulfonic acid (6:2 FTS)3.745427.2407.2
11.Perfluorooctanoic acid (PFOA)3.775413.2369.1
12.Perfluoroheptane sulfonic acid (PFHpS)3.805449.280.0
13.Perfluorononanoic acid (PFNA)4.155463.2419.2
14.Perfluorooctane sulfonic acid (PFOS)4.165499.280.0
15.2-Perfluorooctyl ethanoic acid (FOEA)4.26250477.2393.2
16.1H, 1H, 2H, 2H-Perfluorodecane sulfonic acid (8:2 FTS)4.455527.2507.2
17.Perfluorodecanoic acid (PFDA)4.475513.2469.2
18.Perfluorononane sulfonic acid (PFNS)4.485549.280.0
19.N-Methyl perfluorooctane sulfonamidoacetic acid (N-MeFOSAA)4.605570.2419.2
20.Perfluorodecane sulfonic acid (PFDS)4.745599.280.0
21.N-Ethyl perfluorooctane sulfonamidoacetic acid (N-EtFOSAA)4.745584.2419.2
22.Perfluoroundecanoic acid (PFUnA)4.745563.2519.2
23.Perfluorooctane sulfonamide (FOSA)4.805498.278.0
24.2-Perfluorodecyl ethanoic acid (FDEA)4.84250577.2493.2
25.Perfluorododecanoic acid (PFDoA)4.985613.2569.2
26.1H, 1H, 2H, 2H-Perfluorododecane sulfonic acid (10:2 FTS)4.995627.2607.2
27.Perfluorotridecanoic acid (PFTrDA)5.195663.2619.2
28.N-Methyl perfluorooctane sulfonamide (N-MeFOSA)5.325512.2169.1
29.N-Methyl perfluorooctane sulfonamidoethanol (N-MeFOSE)5.3250616.359.1
30.Perfluorotetradecanoic acid (PFTeDA)5.375713.2669.2
31.N-Ethyl perfluorooctane sulfonamide (N-EtFOSA)5.485526.2169.1
32.N-Ethyl perfluorooctane sulfonamidoethanol (N-EtFOSE)5.4850630.359.1
33.Perfluorohexadecanoic acid (PFHxDA)5.675813.3769.3
34.Perfluorooctadecanoic acid (PFODA)5.895913.3869.3
image
LC_EV0551
ColumnRaptor C18 (cat.# 9304A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Diluent:80:20 Water:methanol
Conc.:5-250 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 5 mM ammonium acetate
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.48020
6.000.4595
6.500.4595
6.510.48020
8.000.48020
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

perfluorinated compound

,

PFC

,

perfluoroalkyl

,

fluorotelomer

,

PFOA

,

PFOS

,

PFAS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us