share:

This chromatogram may contain analytes that are synonyms of your search terms.

Perfluorinated Alkyl Acids (PFBS, PFOA, PFOS, and HFPO-DA/GenX) on Raptor C18 5 μm

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Perfluorobutanesulfonic acid (PFBS)1.5610298.879.9
2.Hexafluoropropylene oxide-dimer acid (HFPO-DA/GenX)2.0210328.9284.9
3.Perfluorooctanoic acid (PFOA)2.475413.1368.9
4.Perfluorooctanesulfonic acid (PFOS)2.6510498.880.0
image
LC_EV0553
ColumnRaptor C18 (cat.# 9304552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Diluent:75:25 Water:methanol
Conc.:5-10 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 5 mM ammonium acetate
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.47525
3.000.4595
3.010.47525
4.000.47525
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PFBS

,

HFPO-DA

,

GenX

,

PFOA

,

PFOS

,

PFAS

,

PFC

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us