share:

EPA Method 537.1 on Raptor C18

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Perfluorobutanesulfonic acid (PFBS)2.0610298.879.9
2.Perfluoro-n-[1,2-13C2]hexanoic acid (13C2-PFHxA)3.035314.9270.0
3.Perfluorohexanoic acid (PFHxA)3.045312.9268.7
4.Tetrafluoro-2-heptafluoropropoxy-13C3-propanoic acid (13C3-HFPO-DA)3.365286.8168.7
5.Hexafluoropropylene oxide dimer acid (HFPO-DA)3.385285.0168.9
6.Perfluoroheptanoic acid (PFHpA)4.095362.8318.8
7.Perfluorohexanesulfonic acid (PFHxS)4.2210398.879.9
8.4,8-Dioxa-3H-perfluorononanoic acid (ADONA)4.245376.9250.7
9.Perfluoro-[1,2-13C2]octanoic acid (13C2-PFOA)4.885415.0370.0
10.Perfluorooctanoic acid (PFOA)4.905413.1368.9
11.Perfluorononanoic acid (PFNA)5.545462.9418.9
12.Perfluoro-1-[1,2,3,4-13C4]octanesulfonic acid (13C4-PFOS)5.5710503.080.0
13.Perfluorooctanesulfonic acid (PFOS)5.5810498.980.0
14.9-Chlorohexadecafluoro-3-oxanone-1-sulfonic acid (9Cl-PF3ONS)5.885530.8350.7
15.Perfluoro-n-[1,2-13C2]decanoic acid (13C2-PFDA)6.085514.9469.9
16.Perfluorodecanoic acid (PFDA)6.085512.9469.0
17.N-deuteriomethylperfluoro-1-octanesulfonamidoacetic acid (d3-NMeFOSAA)6.2810572.9418.8
18.N-methyl perfluorooctanesulfonamidoacetic acid (NMeFOSAA)6.3010569.8418.8
19.N-deuterioethylperfluoro-1-octanesulfonamidoacetic acid (d5-NEtFOSAA)6.5110588.9418.8
20.N-ethyl perfluorooctanesulfonamidoacetic acid (NEtFOSAA)6.5210583.8418.8
21.Perfluoroundecanoic acid (PFUnA)6.555562.9518.8
22.11-chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (11Cl-PF3OUdS)6.775630.7451.0
23.Perfluorododecanoic acid (PFDoA)6.955612.7568.9
24.Perfluorotridecanoic acid (PFTrDA)7.305662.7618.8
25.Perfluorotetradecanoic acid (PFTA)7.605712.7668.7
image
LC_EV0560
ColumnRaptor C18 (cat.# 9304A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Diluent:96:4 Methanol:water
Conc.:5-10 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:Water, 5 mM ammonium acetate
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.47030
8.000.41090
8.010.47030
10.00.47030
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EPA537

,

PFAS

,

Raptor

,

PFOA

,

PFOS

,

GenX

,

ADONA

,

HFPO-DA

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us