share:

This chromatogram may contain analytes that are synonyms of your search terms.

Glyphosate in Bottled Drinking Water at 100 ppt on Raptor Polar X by Large Volume Injection Using LC-MS/MS

PeakstR (min)Conc.
(ng/L)
Precursor IonProduct Ion 1Product Ion 2
1.Glyphosate6.13100168.1062.9578.95
image
LC_EV0571
ColumnRaptor Polar X (cat.# 9311A32)
Dimensions:30 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:30 °C
SampleGlyphosate (cat.# 32426)
Diluent:Direct injection of bottled drinking water fortified at 100 ppt with glyphosate.
Inj. Vol.:500 µL
Mobile Phase
A:Water, 0.5% formic acid
B:Acetonitrile, 0.5% formic acid
Time (min)Flow (mL/min)%A%B
0.000.5595
1.00.5595
3.00.59010
9.00.59010
9.010.5595
10.50.5595
DetectorMS/MS
Ion Mode:ESI-
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

glyphosate

,

large

,

volume

,

drinking

,

water

,

injection

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.