share:

This chromatogram may contain analytes that are synonyms of your search terms.

Pesticides on Ultra Aqueous C18

PeakstR (min)Transition 1Transition 2
1.Acrinathrin9.09540 -> 372540 -> 299.9
2.Chlorfluazuron9.24539.9 -> 356.8539.9 -> 519.7
3.Clothianidin3.63249.9 -> 58247.9 -> 59
4.Diflubenzuron7.68309 -> 156.1310 -> 288.9
5.Diuron6.78230.9 -> 185.8230.9 -> 149.8
6.Fluazinam7.99462.5 -> 415.8462.5 -> 397.9
7.Fludioxonil6.93246.9 -> 179.9246.9 -> 125.9
8.Forchlorfenuron6.44246 -> 126.9246 -> 91.2
9.Hexaflumuron8.45459 -> 438.8459 -> 174.9
10.Imibenconazol8.82409 -> 250.9411 -> 253
11.Lufenuron8.87509 -> 325.9509 -> 175
12.Metaflumizon8.5505.1 -> 301.9505.1 -> 140.9
13.Metamitron3.97201 -> 184.8201 -> 116.9
14.Methoxyfenozid6.98366.9 -> 104.9366.9 -> 148.9
15.Nitenpyram2.59269.2 -> 221.6269.2 -> 100.8
16.Novaluron8.42491.1 -> 470.7493.1 -> 472.7
17.Profoxydim7.57464.3 -> 277.9464.3 -> 126.8
18.Propoxycarbazon3.09396.9 -> 156396.9 -> 112.9
19.Prothioconazol7.16342 -> 100.1343.9 -> 99.9
20.Tebufenozid7.46350.9 -> 149350.9 -> 105
21.Teflubenzuron8.85378.6 -> 338.8378.6 -> 195.9
22.Tepraloxydim4.57340 -> 248340 -> 220.1
23.Terbacil5.38214.9 -> 158.9216.9 -> 160.9
24.Tralkoxydim6.28328.1 -> 253.8328.1 -> 281.8
25.Triadimefon7.01292 -> 67.9292 -> 234.9
26.Triflumuron8356.9 -> 153.8358.9 -> 155.9
image
LC_FF0490A
ColumnUltra Aqueous C18 (cat.# 9178312)
Dimensions:100 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:35 °C
Samplemulticomponent pesticide standard
Diluent:water
Conc.:1 ppb each pesticide
Inj. Vol.:10 µL
Mobile Phase
A:10 mM NH4OAc in water
B:10 mM NH4OAc in methanol
Time (min)%B
0.0020
8.090
12.0100
14.8100
14.920
Flow:0.5 mL/min
DetectorApplied Biosystems/MDS Sciex LC/MS/MS
Model #:4000 QTRAP
Ion Source:TurboIonSpray®
Ion Mode:ESI-
Ion Spray Voltage:-4.2 kV
Gas 1:50 psi (344.7 kPa)
Gas 2:60 psi (413.7 kPa)
Source Temp.:600 °C
InstrumentShimadzu UFLCXR
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

pesticides

,

pesticide residues

,

residue

,

LCMSMS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us