share:

Aflatoxins on Raptor ARC-18 by LC-MS/MS

PeakstR (min)Q1Q3 QuantifierQ3 Qualifier
1.Aflatoxin G21.07331.1245.1189.1
2.Aflatoxin G11.13329.0243.1200.1
3.Aflatoxin B21.23315.0259.1287.1
4.Aflatoxin B11.29313.0285.1241.1
image
LC_FF0538
ColumnRaptor ARC-18 (cat.# 9314A5E)
Dimensions:50 mm x 3.0 mm ID
Particle Size:2.7 µm
Temp.:45 °C
Sample
Diluent:Acetonitrile:water (50:50)
Conc.:100 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:5 mM Ammonium formate + 0.1% formic acid in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.7006535
1.500.700595
1.510.7006535
2.500.7006535
DetectorApplied Biosystems/MDS Sciex LC-MS/MS
Ion Source:TurboIonSpray®
Ion Mode:ESI+
InstrumentShimadzu UFLCXR