share:

This chromatogram may contain analytes that are synonyms of your search terms.

Aflatoxins on Raptor ARC-18 by LC-MS/MS

PeakstR (min)Q1Q3 QuantifierQ3 Qualifier
1.Aflatoxin G21.07331.1245.1189.1
2.Aflatoxin G11.13329.0243.1200.1
3.Aflatoxin B21.23315.0259.1287.1
4.Aflatoxin B11.29313.0285.1241.1
image
LC_FF0538
ColumnRaptor ARC-18 (cat.# 9314A5E)
Dimensions:50 mm x 3.0 mm ID
Particle Size:2.7 µm
Temp.:45 °C
Sample
Diluent:Acetonitrile:water (50:50)
Conc.:100 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:5 mM Ammonium formate + 0.1% formic acid in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.7006535
1.500.700595
1.510.7006535
2.500.7006535
DetectorApplied Biosystems/MDS Sciex LC-MS/MS
Ion Source:TurboIonSpray®
Ion Mode:ESI+
InstrumentShimadzu UFLCXR
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Aflatoxins

,

Raptor

,

ARC-18

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us