share:

This chromatogram may contain analytes that are synonyms of your search terms.

Food Adulterants on Raptor HILIC-Si

PeakstR (min)Precursor IonProduct IonProduct IonPolarity
1.Cyanuric acid0.47127.884.942.1-
2.Cyromazine0.52167.068.285.1+
3.Melamine0.89127.285.068.3+
4.Ammelide2.18129.186.170.2+
5.Ammeline2.97128.286.269.1+
image
LC_FS0515
ColumnRaptor HILIC-Si (cat.# 9310A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:30 °C
Sample
Diluent:5:95 Water:acetonitrile, 10 mM ammonium formate, 0.1% formic acid
Conc.:25 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 10 mM ammonium formate, 0.1% formic acid
B:5:95 Water:acetonitrile, 10 mM ammonium formate, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.60100
0.500.60100
3.500.6595
3.510.60100
8.000.60100
DetectorMS/MS
Ion Mode:ESI+/ESI-
Mode:Scheduled MRM
InstrumentHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Adulterants

,

Melamine

,

HILIC

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us