share:

Bisphenols on Raptor Biphenyl 1.8 μm

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonProduct Ion
1.Bisphenol S0.845.00249.2108.192.1
2.Bisphenol F1.62350199.393.1105.1
3.Bisphenol E2.06100213.3198.3197.4
4.Bisphenol A2.50100227.3212.3133.1
5.Bisphenol AF2.712.00335.2265.3177.3
6.Bisphenol B3.13100241.3212.4211.3
7.Bisphenol C3.43350255.3240.4147.3
8.Bisphenol AP3.9825.0289.3274.3273.3
9.Bisphenol Z4.25250267.2173.4145.2
10.Bisphenol G4.72250311.2295.4296.4
11.Bisphenol FL4.9050.0348.8256.2-
12.Bisphenol BP5.1450.0351.2273.3274.3
13.Bisphenol M5.3915.0345.2330.3251.4
14.Bisphenol P5.6750.0345.2330.4315.3
15.Bisphenol PH6.11350379.2209.4364.4
image
LC_FS0518
ColumnRaptor Biphenyl (cat.# 9309252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Pore Size:90 Å
Temp.:25 °C
Sample
Diluent:75:25 Water:methanol
Conc.:2.00-350 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:Water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.455050
6.500.451090
6.510.455050
8.000.455050
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Bisphenol

,

BPA

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us