Acrylamide Extracted from Potato Chips on Allure Acrylamide

PeakstR (min)Conc.
(ng/mL)
PrecursorProduct
1.Acrylamide-d3 (IS)1.73920075.158.1
2.Acrylamide1.777Endogenous72.155.1
image
LC_FS0530
ColumnAllure Acrylamide (cat.# 9167552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:60 Å
Guard Column:Allure Acrylamide 10 mm, 2.1 mm ID, 5 µm (cat.# 916750212)
Temp.:22 °C
Sample
Diluent:Water
Inj. Vol.:10 µL
Mobile Phase
A:0.001% Formic acid in water
B:0.001% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.41000
1.000.41000
2.000.41090
2.010.41000
5.500.41000
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC
NotesExtracted per EN 16618:2015

Weighed 2.0 g of homogenized potato chips into a 50 mL centrifuge tube. Added 40 mL water followed by the addition of internal standard. Shook by hand for 30 sec, by vortexer for 15 sec, and then on a mechanical shaker for 60 min set to maximum sample extraction agitation. Centrifuged in a refrigerated centrifuge at 10 °C, 3600 x g for 20 min. Removed the aqueous layer after centrifugation, taking care to avoid the top, fatty layer, or the solids at the bottom of the tube. Placed the aqueous extract in an appropriate container.

For cleanup, the first SPE cartridge (multimode SPE column with nonpolar, SAX, and SCX properties, 1000 mg/6 mL) was conditioned with 3 mL methanol and x2, 6 mL aliquots of water. Passed 10 mL of the aqueous extract through the column and collected eluate. For the next cleanup step, the second SPE cartridge (crosslinked polystyrene/poly-DVB SPE column, 500 mg/6 mL) was conditioned with 5 mL methanol and 5 mL water. Passed the eluate from the previous step entirely through the column. Rinsed the loaded cartridge once with 4 mL water and discarded the rinsing solvent. Eluted the acrylamide with 2 mL of 60% methanol in water. Collected the sample and transferred into an evaporation tube. Placed the tube in an evaporator at a temperature no higher than 40 °C to remove the methanol. Evaporated until the final volume was 0.5–0.8 mL using a gentle flow of nitrogen. Transferred the final sample into an autosampler vial and analyzed by LC-MS/MS.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Potato

,

Chips

,

Acrylamide

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us