share:

This chromatogram may contain analytes that are synonyms of your search terms.

Analysis of Bisphenol A on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonQuantifer Product IonQualifier Product Ion
1.Bisphenol A (BPA)1.27227.1212.1133.1
image
LC_GN0559
ColumnRaptor Biphenyl (cat.# 9309552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Temp.:30 °C
Sample
Diluent:50:50 Mobile phase A:mobile phase B
Conc.:100 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:5 mM Ammonium acetate in water
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.57525
2.000.50100
2.010.57525
4.000.57525
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentHPLC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us