share:

This chromatogram may contain analytes that are synonyms of your search terms.

Mycotoxins (STD 1 - 2 ng/g) in MCT Oil on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Aflatoxin G2-13C171.340348.3330.3-
2.Aflatoxin G21.344331.2189.3115.2
3.Aflatoxin G1-13C171.576346.3257.3-
4.Aflatoxin G11.579329.2243.2215.3
5.Aflatoxin B2-13C171.707332.3303.3-
6.Aflatoxin B21.711315.3287.2243.3
7.Ochratoxin A1.824404.3239.1358.3
8.Ochratoxin A-13C201.825424.3250.2-
9.Aflatoxin B11.946313.2241.2128.2
10.Aflatoxin B1-13C171.948330.3301.4-
image
LC_GN0585
ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:35 °C
Sample
Diluent:45:55 Water:Methanol
Conc.:2 ng/g
Inj. Vol.:5 µL
Mobile Phase
A:Water, 2 mM ammonium formate, 0.1% formic acid
B:Methanol, 2 mM ammonium formate, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.73565
2.000.71090
2.010.73565
3.000.73565
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesA working calibration solution containing Aflatoxin B1, B2, G1, and G2 and Ochratoxin A was prepared at 50.0 ng/mL in methanol. 0.25 g of MCT oil was spiked to 2 ng/g using 10 µL of the working calibration solution. A working internal standard was prepared using 13C labeled analogs at a concentration of 250 ng/mL in methanol. 10 µL of the working internal standard was aliquoted into the sample followed by vortexing for 10 seconds at 3,000 rpm. 1 mL of 45:55 H2O:MeOH was added to the sample. The sample was vortexed for 30 seconds at 3,000 rpm. The sample was then centrifuged at 3,000 xg for 5 min. at 10 °C. 750 µL of the supernatant was transferred to a conditioned (1 mL 45:55 water:methanol) Resprep bonded reversed phase SPE cartridge (Restek cat.# 26030). The sample was pulled through under vacuum into an autosampler vial for LC-MS/MS analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

mycotoxins

,

cannabis

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us