share:

This chromatogram may contain analytes that are synonyms of your search terms.

Cannabis Chocolate Sample on Raptor ARC-18 1.8 μm by UHPLC-UV

PeakstR (min)
1.Δ9-Tetrahydrocannabinol (Δ9-THC)2.49
image
LC_GN0593
ColumnRaptor ARC-18 (cat.# 931421E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:1.8 µm
Pore Size:90 Å
Guard Column:UltraShield UHPLC precolumn filter (cat.# 25809)
Temp.:30 °C
Sample
Diluent:25:75 Water:methanol
Inj. Vol.:2 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.001.02575
4.001.02575
DetectorUV/Vis @ 228 nm
InstrumentUHPLC
Notes2 g of chocolate was weighed into a 50 mL centrifuge tube. The sample was brought to a total volume of 40 mL with IPA. The sample was then sonicated at 40 °C for 5 minutes followed by gentle mixing by hand. This was repeated for an additional two cycles. The sample was stored in a -20 °C freezer for 30 minutes to allow the lipids to precipitate. The sample was then vortexed for 5 seconds followed by centrifugation at 3000 rpm for 5 minutes. The supernatant was diluted 10-fold in 25:75 water:methanol, vortexed briefly, and filtered using a 0.2 μm Thomson SINGLE StEP standard filter vial (cat.#: 25893) prior to analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

cannabinoids

,

cannabis

,

potency

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us